贵阳市15套检测系统测定血清尿素氮的比对调查Comparison of 15 Sets of Detection System for Determination of Serum Urea Nitrogen in Guiyang
朱丽英,潘卫,李兴,唐筑灵
ZHU Liying,PAN Wei,LI Xing,TANG Zhuling
摘要(Abstract):
目的:了解贵阳市部分医院临床实验室血清尿素氮检测的差异。方法:向贵阳市10家医院临床实验室15套检测系统发放添加不同浓度尿素氮标准物质的冰冻混合人血清样本5支,尿素氮定值为4.7、9.26、13.8、22.06和31.07 mmol/L,测定尿素氮水平,以水平2(含尿素氮标准物9.26 mmol/L)血清样本作为校准品,对15套检测系统进行校准后重新测量尿素氮水平,观察校准前后不同检测系统测定结果的变异(CV)及测定结果与靶值的偏倚。结果:在统一校准前,实验室间各检测系统尿素氮测定结果的总CV为4.83%~6.8%,封闭检测系统测定结果的CV为3.95%~4.62%,非封闭检测系统测定结果的CV为5.23%~8.08%;统一校准后分别为3.18%~4.71%,1.96%~3.4%及3.59%~5.5%,校准前15套系统测定结果的总CV均高于校准后测定结果的总CV,无论是校准前还是校准后非封闭检测系统测定结果的CV均大于封闭检测系统;在统一校准前,各测定系统对4水平样本测定结果与靶值的相对偏倚分别为0.85%~27.23%、-3.55%~12.88、-3.5%~12.19%、-3.12%~11.04%;统一校准后分别为-1.7%~19.36%、-2.75%~10.06%、-3.26%~5.17、-3.54%~8.56%,比校准前减小。结论:各医院检验科各检测系统血清尿素氮测定结果的变异较大,校准后变异明显下降;封闭检测系统测定结果的可比性好,开放检测系统测定结果可比性较差。
Objective: To understand the difference of serum urea nitrogen detection in different laboratories in some hospitals in Guiyang. Methods: Five frozen human serum standard material with urea nitrogen fixed value being 4. 7,9. 26,13. 8,22. 06 and 31. 07 mol / L were distributed to 15 sets of detection system of ten clinical laboratories in Guiyang. Urea nitrogen level was determined. With level2 serum samples( containing urea nitrogen standard 9. 26 mmol / L) being calibrator,15 sets of detection system was calibrated,after which the level of urea nitrogen was re-detected,and the results of variation( CV) and the target value of bias before and after calibration were observed. Result: Before uniform calibration,the total CV was 4. 83% ~ 6. 8%,the CV of closed detection systems results was3. 95% ~ 4. 62% and the CV of unclosed measurement systems results was 5. 23% ~ 8. 08%. After uniform calibration,the total CV was 3. 18% ~ 4. 71%,the CV of closed detection systems results was1. 96% ~ 3. 4% and the CV of unclosed measurement systems results was 3. 59% ~ 5. 5%. The totalCV of the calibration results of the 15 sets of system before uniform calibration was higher than the total CV after the calibration. The results of CV were larger than those of the closed test system before and after calibration. Before uniform calibration,the relative bias of different measurement systems results were as follows: 0. 85% ~ 27. 23%,- 3. 55% ~ 12. 88,- 3. 5% ~ 12. 19% and- 3. 12% ~11. 04%. After calibration,the relative bias of different measurement systems results were as follows:- 1. 7% ~ 19. 36%,- 2. 75% ~ 10. 06%,- 3. 26% ~ 5. 17 and- 3. 54% ~ 8. 56%. Conclusions: The test results of urea nitrogen from different laboratories vary significantly,especially for the results of unclosed measurement systems. The CV of each measurement system results decreased significantly when the measurement systems were calibrated. The results of the closed detection system were better than that of the open system.
关键词(KeyWords):
血尿素氮;实验室,医院;校准;偏倚
blood urea nitrogen;laboratories,hospital;calibration;bias
基金项目(Foundation): 贵州省科技计划项目[黔科合计省合(2013)7003]
作者(Author):
朱丽英,潘卫,李兴,唐筑灵
ZHU Liying,PAN Wei,LI Xing,TANG Zhuling
DOI: 10.19367/j.cnki.1000-2707.2016.07.017
参考文献(References):
- [1]卢志芳,杨先凯,李凤侠.室间不同生化检测系统结果可比性探讨[J].实验与检验医学,2014(1):34-36.
- [2]王丽,权翠侠,牛璐璐,等.新鲜血清赋值传递提高不同生化检测系统检验结果的可比性[J].临床检验杂志,2008(4):294-296.
- [3]朱松山,殷君,陆小彩,等.不同生化检测系统间结果可比性的探讨[J].重庆医学,2013(10):1150-1151.
- [4]张蕾.检验结果互认工作的难点及对策探讨[J].卫生职业教育,2014(5):94-95.
- [5]Huang X,Li S,Yang X,et al.Establishing reference values for blood urea nitrogen and serum creatinine in Chinese Han ethnic adult men[J].Clin Lab,2014(7):1123-1128.
- [6]丛玉隆.临床实验室管理[M].北京:中国医药科技出版社,2004:48-50.
- [7]邱玲,程歆,刘荔,等.多台生化分析仪多项目同时进行比对的实验设计及应用[J].中华检验医学杂志,2007(9):1001-1004.
- [8]张宏杰,高峰,尹家庆,等.3台不同生化分析仪系统检测血尿素氮、肌酐的比对及偏倚分析[J].国际检验医学杂志,2013(B12):69-71.
文章评论(Comment):
|
||||||||||||||||||
|
||||||||||||||||||