应用极差分析法比对血细胞分析仪的检测结果The Comparative Analysis on the Results among the Hematology Analyzers by Range Analysis Method
梁凌云,徐华,李彬,陈佳
LIANG Lingyun,XU Hua,LI Bin,CHEN Jia
摘要(Abstract):
目的:应用极差分析法对4台检测原理相同的Sysmex全自动血细胞分析仪进行比对,探讨检测结果的可比性。方法:依据极差检验可比性方案,选择合适浓度标本,在4台编号分别为J-724-LJ、J-727-LJ、J-729-LJ、J-734-LJ的血细胞分析仪上检测WBC、RBC、Hb、HCT、MCV、MCH及MCHC、PLT,记录测定结果,计算极差,进行比对。结果:Sysmex全自动血细胞分析仪J-734-LJ的RBC、HCT、MCH、MCHC检测结果偏高,超出允许范围;排除来自仪器J-734-LJ的极端值后重新计算比对,J-724-LJ、J-727-LJ、J-729-LJ三台仪器的8大项目检测结果的极差均小于临界差值,具有可比性,检测结果差异无统计学意义。结论:极差分析法可用于检测原理相同的血细胞分析仪检测结果的比对。
Objective: Application of range analysis method to four Sysmex automated hematology analyzers in order to investigate the comparability. Methods : According to range analysis method,appropriate concentrations of specimens were chosen,tested on four automated hematology analyzer,the determination results were recorded. Results: RBC,HCT,MCH,MCHC test results of Sysmex automated hematology analyzer J-734-LJ were high,beyond the permitted range. After excluding the extreme value of J-734-L,range analysis test results of eight items of J-724-LJ,J-727-LJ,J-729-LJ were less than the critical difference,and the results were comparable,there was no significant difference among instruments. Conclusion: Range analysis method is suitable for detecting comparability among hematology analyzers with the same principle.
关键词(KeyWords):
极差分析;血细胞分析仪;质量控制;血细胞计数
range analysis;hematology analyzer;quality control;blood cell count
基金项目(Foundation):
作者(Author):
梁凌云,徐华,李彬,陈佳
LIANG Lingyun,XU Hua,LI Bin,CHEN Jia
DOI: 10.19367/j.cnki.1000-2707.2013.06.016
参考文献(References):
- [1]林颜玉,吴媛.不同型号血液分析仪对同一标本WBC、Hb和PLT检测结果的比较[J].白求恩军医学院学报,2011(3):227-228.
- [2]王治国.临床检验方法确认与性能验证[M].北京:人民卫生出版社,2009:319-339.
- [3]River PA,Dobalian A.A review and analysis of the clinical laboratory improvement amendment of 1988:compliance plans and enforcement policy[J].Health Care Manage Rev,2005(2):93-102.
- [4]高峰林.如何保证第三方实验室检测结果的准确性[J].分析试验室,2008(27):443-445.
- [5]黄保荣,薛莲,王金松.运用极差检验对两个监测系统上糖化血红蛋白结果进行可比性验证[J].国际检验医学杂志,2012(13):1641-1643.
- [6]周艳.极差法比对和EP9-A2指南在不同仪器系统间可比性的联合应用[J].长江大学学报(自科版),2013(24):87-92.
- [7]彭黎明,府伟灵,杨青,等.生物学变异对正常人血小板数影响的研究[J].血栓与止血学,2005(1):29-32.